Agilent Technologies B1500A Betriebsanweisung Seite 501

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 628
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 500
Agilent B1500 Users Guide, Edition 7 9-27
If You Have a Problem
Error Codes
105073 Hold time for I/V-t sampling must be between A s and B s.
105074 Multiplying the number of monitor channels by No of Samples must be
between A and
B.
105075 Invalid value A specified for I/V
-t sampling Lin/Log setting.
Valid option(s): B
105076 Base hold time for I/V-t sampling must be between A s and B s.
105077 In
valid value A specified
for I/V-t sampling output sequence.
Valid option(s): B
105078 Sampling interval for I/V-t A sampling must be between B s a
nd C s.
105079 Hold time for I/V-t A sampling must be betw
een B s and C s.
105080 Hold time for I/V-t A sampling with interval of B s or longer must be
between C s a
nd D s.
105081 Hold time for I/V-t A sampling with interval shorter than B s must be
between C s a
nd D s.
105083 The number of VAR1 steps multiplied by the number of VAR2 steps
must be between A and
B in I/V Sweep, I/V List Sweep and Multi
Channel I/V Sweep.
105091 List length of VAR1 sweep must be between A and B for the I/
V List
Sweep measurement. Length C is not applicable.
105092 List length of VAR2 sweep must be between A and B for the I/
V List
Sweep measurement. Length C is not applicable.
105093 List length of delay time must be between A and B for the I/V List
Sweep measur
ement. Length C is not applicable.
105094 Time until the first step measurement, (hold time + delay time), must be
between A and B s fo
r the I/V List Sweep measurement using SMU
pulse.
105095 A does not support current measurement.
105801 Inva
lid value A specified fo
r Output Value after Measurement in the
Advanced Setup dialog box. Valid option(s): B
105802 Invalid value A specified fo
r Bias Hold after Measurement in the
Advanced Setup dialog box. Valid option(s): B
105803 Bias hold function execution error. All channel outputs or voltage
compliance must be at ±42 V or l
ess to control the output switch or the
series resistor.
Seitenansicht 500
1 2 ... 496 497 498 499 500 501 502 503 504 505 506 ... 627 628

Kommentare zu diesen Handbüchern