7- 54 Agilent B1500 User’s Guide, Edition 7
Function Details
SMU Compliance
Power Compliance
In addition to V compliance or I compliance, you can set power compliance for the
staircase sweep source.
If you specify power compliance, the source/monitor unit (SMU) changes the
current or voltage compliance v
alue every sweep step. The value is the lower value
of either the specified compliance value or the compliance value given by the
following formula. See Figure 7-25.
Current compliance = Specified power compliance valu
e / Step voltage
Voltage compliance = Specified power compliance value / Step
current
Figure 7-25 Power Compliance Output Area
Available values The following values are available for the power compliance:
MPSMU/HRSMU 1 mW to 2 W
HPSMU 1 mW to 20 W
Conditions after
compliance
When a chan
nel reaches compliance, the staircase sweep source automatically stops
the sweep and forces the start value. After reaching the compliance, the
measurement data will be “NaN”.
9ROWDJH
&XUUHQW
$FWXDOFXUUHQWFRPSOLDQFH
6SHFLILHGFXUUHQWFRPSOLDQFHYDOXH
)RUYROWDJHVZHHS )RUFXUUHQWVZHHS
6SHFLILHGSRZHUFRPSOLDQFH
&XUUHQW
9ROWDJH
$FWXDOYROWDJHFRPSOLDQFH
6SHFLILHGYROWDJHFRPSOLDQFHYDOXH
6SHFLILHGSRZHUFRPSOLDQFH