Agilent Technologies B1500A Betriebsanweisung Seite 174

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 628
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 173
3- 44 Agilent B1500 User’s Guide, Edition 7
Installation
Connecting Measurement Devices
Guarding
Guarding reduces the leakage current between the measurement points
and
instrument. This is important when you measure low current.
The following figure shows the theory of guarding. The buffer amplifier (×1) keeps
the potenti
al of the guard conductor at the same potential as the force conductor, so
current does not flow between the force and guard conductors. Therefore, the
current measured by SMU is same as current at measurement point because no
current is leaked.
$
%XIIHU
*XDUG
)RUFH
'87
Seitenansicht 173
1 2 ... 169 170 171 172 173 174 175 176 177 178 179 ... 627 628

Kommentare zu diesen Handbüchern

Keine Kommentare