Seitenansicht 229
4- 42 Agilent B1500 User’s Guide, Edition 7
Using EasyEXPERT
Configuration
Start Diagnosis Starts diagnostics for the selected items. To perform diagnostics, specify the items to
perform and click this button. A dialog box appears. Follow the dialog box to
perform the diagnostics. Some items need your judgement, pass or fail. Also you
will need to connect an adapter, press keys, check LEDs, and so on.
Modules
The Modules tab screen is used to perform self-test of modules.
Slot Configuration This area lists Slot, Module Type, Name, and Status.
Slot Slot number where the module is installed. To
select, check the
left check box.
Module Type Module model number
Name Module name
Status Self-test status, Passed, Failed, Recovered, or ---- (s
elf-te
st
has not been performed)
SCUU Specifies if the module is connected to the
SCUU.
Notes Error information when the module self-test failed.
Select All Selects all modules for self-test. This checks the check box for all modules.
Unselect All Unselects all modules. This unchecks the box for all modules.
Start Self Test Starts self-test for the selected modules.
Recover Module This button is for service use only. Makes the failed modules respond again. This
b
utton do
es not solve the problem that causes the Failed status.
Abort Aborts self-test.
Status LED Enables (ON) or disables (OFF) the status i
ndicator (LED) of
the ASU (Atto Sense
and Switch Unit) and the SCUU (SMU CMU Unify Unit).