HIGH VOLTAGE SHOCK HAZARD
Agilent B1500 can force dangerous voltages (200 V for HPSMU, and 100 V for
MPSMU/HRSMU) at t
he force, guard, and sense terminals. To prevent electric
shock hazard, the following safety precautions must be observed during the use of
Agilent B1500.
Use a three-conductor AC power cable to connect cabinet (if used) and Agilen
t
B1500 to an electric ground (safety ground).
If you do not use Agilent 16442 Test Fixture, make sure to connect the Interlock
terminal to a swit
ch that turns off when the shielding box access door is opened.
Confirm periodically that interlock function works normally.
Before touching the connections of the force, guard, and sense terminals, turn
Agilent
B1500 off and discharge any capacitors whenever possible. If you do
not turn Agilent B1500 off, complete “all” of the following items, regardless of
any Agilent B1500 settings.
Terminate measurement by pressing Stop key, confirm that the Measurement
status in
dicator is not lit.
Confirm that the
High Voltage indicator is not lit.
Open the shielding box access door (open the Interlock terminal).
Discharge any capacitors if the capacitance is connected to an SMU.
Warn workers in the vicinity of Agilent B1500 about dangerous conditions.