4- 40 Agilent B1500 User’s Guide, Edition 7
Using EasyEXPERT
Calibration
Integration Time This area defines the integration time used for measuring phase compensation data
or open/short/load correction data. The number of averaging samples
(Mode=AUTO) or the averaging time (Mode=PLC) is set.
Mode A/D converter operation mode, AUTO or PLC
Factor Factor for the initial value. See “Mode, Factor” on page 7-20.
Reference
Standard
Values
The following entry fields are available to set the calibration values or the reference
values of the open/short/load standard. If you use the standard, enter the values to
the following entry fields.
OPEN Reference: Capacitance and Conductance
If you do not use the reference, enter 0 in the both fields.
SHORT Reference: Inductance and Resistance
If you do not use the reference, enter 0 in the both fields.
LOAD Reference: Inductance and Resistance