Agilent B1500 Users Guide, Edition 7 9-25
If You Have a Problem
Error Codes
105031 The product of voltage source value and current compliance value of A
exceeds its output power range. Please set the current compliance value
between B A and C A.
105032 The product of current source value and voltage compliance value of A
exceeds its output po
wer range. Please set the voltage compliance value
between B V and C V.
105033 The polarity of peak value(s) and the base value of SMU current pulse
must be th
e same.
105034 The start value and stop value of SMU log sweep must be non-zero and
have same pol
arity for primary sweep source (and synchronous sweep
source).
105035 Invalid value (A) was specified fo
r SMU I/V Sweep stop condition.
Valid option(s): B
A is not available for the setting of the sweep abort condition. Select
one from the available ch
oices B. If you use a formula, the calculation
result must show the available value.
105036 The hold time for SMU I/V Sweep must be between A s and B s.
105037 The delay
time for SMU I/V Sweep must be between A s and B s.
1050
41 Invalid value (B) was specified for current
monitor ranging mode of A.
Valid option(s): C
105042 Invalid value (B) was specified fo
r current monitor range of A. Valid
option(s): C
105043 Invalid value (B) was specified fo
r current monitor ranging rule of A.
Valid option(s): C
105044 Invalid value (B) was specified for volt
age monitor ranging mode of A.
Valid option(s): C
105045 Invalid value (B) was specified fo
r voltage monitor range of A. Valid
option(s): C
105046 Invalid value (A) was specif
ied for the integral mode of
High-Resolution ADC. Valid option(s): B
105047 The coefficient for High Resolution ADC A integr
al mode must be
between B and C.
105048 Invalid value (A) was specif
ied for the averaging mode of High-Speed
ADC. Valid option(s): B
Seitenansicht 498