Agilent B1500 User’s Guide, Edition 7 4- 47
Using EasyEXPERT
Configuration
Cable Specifies the model number of the cable from the
B2200A/ B2201A output to the DUT I/F. The selections
are the 16494A/B/C-001 (1.5 m) / 002 (3 m) / 005 (4 m).
This mode cannot perform the capacitance compensation fo
r the cables to
positioner/manipulator or the probe card.
• User Compensation Data File
Select this radio button and specify the compensation data file to perform the
capacitance compensa
tion for the path from the B2200A/B2201A input
ports to the end of positioner/manipulator/probe card. See “To Create
Compensation Data File” on page 3-50. The data after compensation is
displayed and saved.
• E5250A (option)
The following setup is available for the E5250A hardware profile.
• User Compensation Data File
Select this radio button and specify the compensation data file to perform the
capacitance comp
ensation for the path from the E5250A input ports to the
end of positioner/manipulator/probe card. See “To Create Compensation
Data File” on page 3-50. The data after compensation is displayed and
saved.
Clear Clears the User Compensation Data File field.
Triaxial Cable Enter the length of the 16494A triaxial cable
connected between the E5
250A outputs to the
connector plate.
User Triaxial Cable Enter the length of the triaxial cable between
the connector plate
and the DUT. Enter 0 if
the triaxial cable is not used.
User Coaxial Cable Enter the length of the coaxial cable between
the connector plate
and the DUT. Enter 0 if
the coaxial cable is not used.
Close Closes the Extended Configuration dialog box.