Agilent Technologies B1500A Betriebsanweisung Seite 172

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 628
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 171
3- 42 Agilent B1500 User’s Guide, Edition 7
Installation
Connecting Measurement Devices
Connections for High Current Measurements
When you force or measure a large current, you may want to use a Kelvin (4-wire)
connection to eliminate the residual resistance effects of test leads and contacts. For
example, you can use the following connections as Kelvin connections on the test
fixture.
Examples: Kelvin Connection
To cancel the effects of the residual resistance, test leads must be connected as close
as possible to
the DUT.
6089 $0D[
6HQVH)RUFH
*XDUG *XDUG
6HQVH)RUFH
*XDUG *XDUG
6HQVH)RUFH
*XDUG *XDUG
6HOHFWHG
*DWH
6RXUFH
'UDLQ
6XEVWUDWH
Seitenansicht 171
1 2 ... 167 168 169 170 171 172 173 174 175 176 177 ... 627 628

Kommentare zu diesen Handbüchern

Keine Kommentare