Agilent B1500 User’s Guide, Edition 7 2- 63
Introduction
Specifications
Table 2-31 Measurement functions, voltage forcing, voltage measurement, and current
measurement
Table 2-32 Voltage measurement accuracy, resolution, and noise
Table 2-33 Current measurement accuracy and resolution
Mo
de Function
V force
ranges
V measure
ranges
I measure
ranges
Fast
IV V force/I me
asure,
V force/V me
asure
-3 V
to
+3 V,
-5 V to +5 V,
-10 V to 0 V,
0 V to +10 V
-5 V
to +5 V,
-10 V to
+10 V
1 μA, 10 μA,
100 μA,
1 mA, 10 mA
PG V force/V me
asure
-3 V to +3 V,
-5 V to +5 V
-5 V t
o +5 V -
DC V force/I me
asure,
V force/V me
asure
-3 V
to
+3 V,
-5 V to +5 V,
-10 V to 0 V,
0 V to +10 V
-5 V
to +5 V,
-10 V to
+10 V
1 μA, 10 μA,
100 μA,
1 mA, 10 mA
SMU pass
through
Measurement
us
ing SMU
Maximum
± 25 V
- Maximum
± 100 mA
Accuracy ± 0.1 % of reading ± 0.1 % of range
(independent of
the range or mode)
Resolution
a
a. Can vary at most 5 % based on the result of calibration
680 μV (-5 V to +5 V
range), 1.4 mV (-10 V to
+10 V)
Noise
4 mV
rms
(-5 V to +5 V range)
b
b. Into an open load. Without averaging.
Accuracy ± 0.1 % of reading ± 0.1 % of range
(independent of
the range or mode)
Resolution
a
a. Display resolution. Can vary at most 5% based on the results of calibra-
tion.
0.014 % of range