2- 66 Agilent B1500 User’s Guide, Edition 7
Introduction
Specifications
Measurement mode details
Agilent B1500A supports the following measurement modes:
Staircase sweep
Multi-channel sweep (EasyEXPERT does not support VAR1’)
Pulsed sweep
Staircase sweep with pulsed bias
•IV sampling
High speed IV sampling
CV sweep
C-t sampling
C-f sweep
List sweep
Linear search (supported by FLEX command)
Binary search (supported by FLEX command)
Each SMU can be set to VAR1 (primary sweep), VAR2 (secondary sweep), VAR1’
(s
ynchronous
sweep), or CONST (constant voltage/current source).
•VAR1
Primary sweep controls the staircase (DC or puls
ed) voltage or current sweep.
Maximum number of VAR1 steps: N
1
= 1001
•VAR2
Subordinate linear staircase or linear pulsed sweep
. After primary sweep is
completed, the VAR2 unit output value is changed.
Maximum number of VAR2 steps: N
2
= 1001 (condition: 1 N
1
× N
2
128128)
•VAR1
Staircase or pulse sweep synchronized with the VAR1 sweep. Sweep is made
with a user specified r
atio and offset values.
VAR1 = a × VAR1 + b
“a” is the u
ser specified ratio, and “bis the use
r specified offset value.