7- 40 Agilent B1500 User’s Guide, Edition 7
Function Details
Bias Hold Function
Bias Hold Function
This function is used to keep source output after measurement. Source modules
apply the specified bias between measurements in a quick test or application test
that defines some classic test setups, or a repeat measurement. And the source
modules change the output value and the unused modules are disconnected when the
next measurement is started.
Output channels
a
nd output value
If th
e bias hold function is OFF, the channel stops the output immediately after
measurement.
If the function is ON, the channel continues the output as shown below.
I/V Sweep measurement:.
VAR1 and VAR1’ output channels: Sweep start or stop value
VAR2 output channel: Sweep stop value
CONST output channels: Output value
SMU pulse output channels: Base value
C-V Sweep measurement:
Bias sweep output channel: Sweep start or stop value
CONST output channels: Output value
I/V-t Sampling measurement:
Constant source output channels: Base or source value
SPGU output:
SPGU keeps the output as shown in Table 7-6.
Period of bias hold From the end of a measurement to t
he end of a si
ngle measurement
From the end of a measurement to the start of the next measurement during a
quick test, appl
ication test, or repeat measurement
Reset of source
out
put
The sour
ce output is reset after a quick test, application test, repeat measurement, or
single measurement. The SMU is set to 0 V output and 0.1 mA compliance. The
MFCMU is set to 0 V DC bias and 0 V oscillator level.