Agilent B1500 User’s Guide, Edition 7 7- 43
Function Details
Current Offset Cancel
Table 7-7 Measurement range, allowable offset value, and module type
If the offset value for a measurement range of a channe
l exceeds the allowable
value, the offset data is not effective in this range and lower for this channel. Then,
if the ranging mode setting of this channel covers this range or lower, the offset
cancel is not performed for any data measured by this channel.
Measurement
range
Allowable offset value Module type
1 nA ±10 pA HR/ASU, HR, MP, HP
100 pA ±1 pA HR/ASU, HR
10 pA ±400 fA HR/ASU, HR
1 pA ±100 fA HR/ASU