5- 38 Agilent B1500 User’s Guide, Edition 7
Classic Test Definition
Direct Control
compenReZ
This built-in function performs compensation on the specified impedance data and
returns the real data for the compensation results.
This function is effective for the CMU measurement data compensation while the
switching ma
trix is used with the CMU to extend the measurement terminals.
Define the compenReZ and compenImZ functions as the user functions on the
Function Setup screen, so that the compensation is performed while the Direct
Control test is being executed. Also enter the formula to calculate the measurement
parameters. An example definition is shown in Figure 5-2. For the equations, see
“About Impedance Measurements” on page 7-21.
Syntax
compenReZ(freq uency,Real,Imaginary)
Parameters frequency Measurement frequency. Hz. Numeric scalar or vector data.
For the C-f measurement data, frequency shoul
d be a vector
variable that stores multiple frequency values.
Real CMU Re(Z) measurement result data. Ω. Numeric vector data.
Real sh
ould be a vector variable that stores the impedance real
data befo
re compensation.
Imaginary CMU Im(Z) measurement result data. Ω. Numeric vector data.
Imag
inary should be a vector variable that stores the impedance
imaginary data bef
ore compensation.
Example Name Rcompen
Definition compenReZ(Freq,ReZ,ImZ)
Rcompen returns the impedance real data for the compensation results.