5- 10 Agilent B1500 User’s Guide, Edition 7
Classic Test Definition
Multi Channel I/V Sweep
Channel Setup
This setup screen is used to define the channel setups and the variables for the
channel output data, measurement data, time stamp data, and index data.
Channel Definition This area defines the channel setups and the v
ariables for th
e channel output data
and the measurement data.
Unit Module used for the source output or measurement
V Name Variable name of voltage output or measurement data
I Name Variable name of current output or measurement data
Mode Output mode. For SMU, V (voltage), I (current), or COMMON
(g
round). For
SPGU, VPULSE (voltage pulse) or ALWG
(arbitrary linear waveform voltage).
Function Output function, VAR1 (primary sweep), VAR2 (secondary
sweep), or CONST (con
stant)
Add SMU Adds a row for the SMU channel setup.
Add SPGU Adds a row for the SPGU channel setup.
Delete Deletes selected channel setup. To select a channel, use the
r
adio but
ton on the left.
Up Moves the selected channel setup upward.
Down Moves the selected channel setup downward.
If multiple output channels are defined, the channels start output sequentially. The
o
rder is to
p to bottom of the channels defined in this screen. The output stops in the
opposite order.
Miscellaneous
Va
riabl
es
This area defines the variables for the time stamp data and the index data.
Time Stamp Name Variable name of the time stamp. Time stamp data indicates
the time at which st
ep measurement was started.
Index Name Variable name of the data index. Integer above 0 is returned.