4- 32 Agilent B1500 User’s Guide, Edition 7
Using EasyEXPERT
Repeat Measurement Setup
Start Procedure Specifies the procedure executed before starting measurement.
Iteration
Procedure Specifies the proced
ure executed every sing
le measurement.
Final Procedure Specifies the procedure executed after the repeat measurement.
If
the procedure requires arguments, enter the va
lues to the Arguments field.
If the procedure specified by Start Procedure or Iteration Pro
cedure creates the
response as shown in Figure 4-2, EasyEXPERT performs the operation following
the response. See Figure 4-1. The parameter status value mus
t be the string True or
False, and the parameter device_id val
ue should be any string. EasyEXPERT
ignores the response of the different format.
Automatically fill
in Dev
ice ID
This check b
ox is available when the Procedure return condition box has been
checked, and is used to enable the device ID automatic input function. This function
enters the Device ID of the test record automatically. The value is given by
device_id which is a parameter of the response created by Start Procedure and
Iteration Procedure.
Repeat Stop
Condit
ion
This area is
used to set the repeat measurement stop condition. The following stop
conditions are available.
• Counter Reaching to: limit
The repeat
measurement will be stopped if Count ≥ limit.
• Pro
cedure return condition
The repeat measurement will be stopped if the response shows status = Tru
e.
Figure 4-2 Procedure Response Expected by EasyEXPERT
• Response (input to EasyEXPERT):
XML format data
status: True (break) or False
device_id: Device ID
• Example:
<Response>
<Break>status</Break>
<Target>device_id</Target>
</Response>
<Response>
<Break>False</Break>
<Target>waf1a:3 1</Target>
</Response>
Kommentare zu diesen Handbüchern