
48 Chapter 2
Basic Operation
Creating and Applying User Flatness Correction
• GPIB interface cable
• adapters and cables, as required
Configure the Power Meter
1. Select SCPI as the remote language for the power meter.
2. Zero and calibrate the power sensor to the power meter.
3. Enter the appropriate power sensor calibration factors into the power meter as
appropriate.
4. Enable the power meter’s cal factor array.
NOTE For operating information on your particular power meter/sensor, refer to its
operating guide.
Connect the Equipment
Connect the equipment as shown in Figure 2-5 on page 49.
NOTE During the process of creating the user flatness correction array, the power
meter is controlled by the signal generator via GPIB. No other controllers are
allowed on the GPIB interface.
Kommentare zu diesen Handbüchern