90 Chapter 3
Basic Digital Operation
Using Waveform Markers
Using Waveform Markers
The ESG provides you with four waveform markers. Waveform markers are used by the ARB
modulation formats (W-CDMA, CDMA2000, AWGN, etc.) and the dual ARB player to mark
specific points on a played waveform. When markers are encountered by the ESG, auxiliary
output signals are generated and routed to the rear panel event connector (EVENT 1, 2, 3, or
4) that corresponds to the marker number. The markers can also be configured to initiate one
or all three of the following functions: RF blanking, ALC hold, trigger alternate amplitude.
ALC hold is automatically activated when RF blanking occurs. So there is no need to select
marker ALC hold for the same marker number that is utilizing RF blanking.
The auxiliary output signals can be constructed as clock signals for synchronizing other
instruments with the played waveform, or used as a trigger signal for starting measurements
at a given point on the waveform.
When a waveform file is generated, a marker file is created that places a marker on the first
sample point of the waveform for markers one and two. However when a waveform file is
downloaded that does not have a waveform file associated with it, the ESG creates a marker
file, but does not place a marker on the first point.
Within the ARB modulation formats, marker points can be modified only while the waveform
is playing. However in the dual ARB player, the waveform does not have to be playing in order
to modify the marker points. You can also edit the marker points for the
AUTOGEN_WAVEFORM file from the dual ARB player while the waveform is playing within
a modulation format.
When working with waveform files in the dual ARB player, the files, called waveform
segments, must reside in volatile memory. For loading waveform segments into volatile
memory, see “Loading Waveform Segments from Non-Volatile Memory” on page 83.
In the following procedures you will be shown how to use markers while working in the dual
ARB player, but the process is the same for all the ARB modulation formats. The terms
marker points and sample points are used interchangeably.
Accessing the Marker Utilities
This procedure shows you the location of the marker parameters. Unless stated differently by
having you press the
Mode hardkey, the procedures following this one will start after entering
the
Marker Utilities softkey menu.
1. Press
Mode > Dual ARB.
2. Press the
Select Waveform softkey, highlight a waveform for play-back, and press Select