Chapter 16 467
Troubleshooting
Basic Signal Generator Operations
search calibration routine for the current RF frequency and amplitude. In this mode, if there
is a change in RF frequency or amplitude, you will need to press
Do Power Search again.
When
Power Search Manual Auto is set to Auto, the calibration routine is executed whenever
the frequency or amplitude of the RF output is changed.
RF Output Power too Low
1. Look for an OFFS or REF indicator in the AMPLITUDE area of the display.
OFFS tells you that an amplitude offset has been set. An amplitude offset changes the value
shown in the AMPLITUDE area of the display but does not affect the output power. The
amplitude displayed is equal to the current power output by the signal generator hardware
plus the value for the offset.
To eliminate the offset, press the following keys:
Amplitude > More (1 of 2) > Ampl Offset > 0 > dB.
REF tells you that the amplitude reference mode is activated. When this mode is on, the
displayed amplitude value is not the output power level. It is the current power output by
the signal generator hardware minus the reference value set by the
Ampl Ref Set softkey.
To exit the reference mode, follow these steps:
a. Press
Amplitude > More (1 of 2).
b. Press
Ampl Ref Off On until Off is highlighted.
You can then reset the output power to the desired level.
2. If you are using the signal generator with an external mixer, see “Signal Loss While
Working with Mixers” on page 463.
3. If you are using the signal generator with a spectrum analyzer, see “Signal Loss While
Working with Spectrum Analyzers” on page 465.
No Modulation at the RF Output
Check the MOD ON/OFF annunciator on the display. If it reads MOD OFF, press Mod On/Off to
toggle the modulation on.
Although you can set up and enable various modulations, the RF carrier is modulated only
when you have also set
Mod On/Off to On.
For digital modulation, make sure that
I/Q Off On is set to On.