192 Chapter 6
BERT Testing
Bit Error Rate Tester–Option UN7
Data Processing
Data Rates
Data rates up to 60 MHz are supported for BER analysis on un-framed or framed PN
sequences. Note that the BER analyzer supports only continuous PN sequences.
Synchronization
Immediately after the trigger event, the DSP for the BER measurement tries to establish
synchronization using the first incoming bit stream.
If the
Bit Delay Off On softkey is set to On, the number of bits specified by the Delayed Bits are
ignored. The synchronization checking is repeated using an error-free bit string, lengthened
by the
Delayed Bits, until synchronization is established.
When the
BERT Resync Off On softkey is set to On, the BER measurements will automatically
be restarted if the intermediate BER measurement result exceeds the value specified by the
BERT Resync Limits.
Special Pattern Ignore Function
The special pattern ignore function is especially useful when performing BER analysis on
radios that generate consecutive 0’s or 1’s data for traffic channels when they fail to detect the
Unique Word or lose synchronization. If 80 consecutive incoming data bits are all either 1’s or
0’s, when the
Spcl Pattern Ignore Off On softkey is set to On, all of the consecutive 0’s or 1’s (and
several bits before and after the consecutive 0’s or 1’s) are ignored. The following figure shows
the operation example of the special pattern ignore function.
Figure 6-28
Pass/Fail Judgement
There are two pass/fail judgement update modes: cycle end and fail hold. With cycle end
selected, either pass or fail judgement is made for the results of each measurement cycle. With
fail hold selected, the fail judgement is retained whenever a failure occurs during one loop of
BER repeat measurements. Fail hold mode allows you to determine when a failure occurs at
least once during an entire cycle of measurements.