60 1100 Series Capillary LC System Manual
4 Basic System Troubleshooting
Dirty solvent inlet filters.
Solvent intake is being
restricted. %Ripple might
also be too high.
Temporarily remove solvent inlet filters to see if
they are the cause of the problem. If so, clean or
replace the solvent inlet filters.
To minimize this problem, prefilter
the mobile phase, and take
precautions against algae formation
in the water.
Dirty EMPV (micro mode
only)
Perform the EMPV cleaning procedure, followed
by the EMPV performance test. Refer to the
Capillary Pump Reference Manual.
Any system component
which is offering a changing
restriction to the pump.
Replace the analytical column.
Replace the filter frit in front (upstream) of the
EMPV.
The vacuum micro-degasser
is off, or has become
defective.
Try another vacuum micro-degasser, or
experiment to determine the performance using
different degasser channels.
If the mobile phase is very sensitive to
gaseousness, use the micro-degasser
continuous mode.
Basic performance problems
in the pump.
Perform the pump Leak Test. Refer to the Capillary Pump
Reference Manual for information on
the leak test.
Unstable column flow and/or system pressure: Possible Causes and Suggested Actions (continued)
Possible Causes Suggested Actions Notes