26 1100 Series Capillary LC System Manual
1 Installing your Capillary LC System
Inject the Check-out Sample
The purpose of the instrument check is to demonstrate that all modules of the
instrument are correctly installed and connected. It is not a test of the
instrument performance.
A single injection of the Agilent Technologies isocratic test sample (Agilent
part number 01080-68704) is made under the method conditions given below:
For system configurations other than those shown above, the method
conditions might need to be altered to produce the desired chromatogram.
Tabl e 4 Method conditions for injecting a test sample
Flow: 15.0 µl/minute
Stoptime: ~7.00 minutes
Solvent A: 30% (HPLC grade Water)
Solvent B: 70% (HPLC grade Acetonitrile)
Wavelength DAD/MWD: Sample: 254/4 nm, Reference: 360/80 nm
Injector Volume: 200 nl
Column Temperature 25.0 °C or ambient
Agilent 1100 Series
Capillary LC Instrument
Degasser
Capillary pump - 20 µl/minute sensor installed
Micro Autosampler
Column Compartment - optional
Detector - DAD with 500 nL flow cell installed
ChemStation
Column: ZORBAX SB C18, 5 µm, 150 x 0.5 mm
Agilent Part No. 5064-8256
Standard: Agilent Part No. 01080-68704
0.15 wt.% dimethylphthalate, 0.15 wt.% diethylphthalate
0.01 wt.% biphenyl, 0.03 wt.% o-terphenyl in methanol
Diluted 1:10 in Acetonitrile