1100 Series Capillary LC System Manual 19
Installing your Capillary LC System 1
Install the Capillary Pump (G1376A)
1 Ensure the line power switch at the front of the capillary pump is off.
2 Place the capillary pump on the bench, positioned to the left of the micro
sampler thermostat module.
3 Connect the power cable to the power socket at the rear of the capillary
pump. Do not connect the power cable to power until you have finished the
hardware installation of all modules in the stack.
4 Connect the 1-meter CAN-bus cable (5181-1519) from one of the CAN-bus
connectors at the rear of the capillary pump to the free CAN-bus connector
at the rear of the micro sampler.
5 Connect the pre-terminated end of the pump-to-sampler capillary
(G1375-87310) to the flow sensor outlet of the pump. Connect the other end
of this capillary to port 1 of the micro sampler injection valve.
6 Connect the 1/8 inch plastic EMPV waste tube to the barbed waste fitting of
the EMPV. Route the waste tube to an appropriate waste container.
7 Connect the large-bore corrugated plastic leak drain tubing to the pump
leak drain fitting. Route the leak drain tubing to an appropriate waste
container.
WARNING
Connect the power cable to power only after you have finished the hardware
installation of all modules in the stack.
NOTE
Carefully route all capillaries so that they are not crushed or broken by module front covers.
Avoid excessive bending. See Chapter 2 for advice on handling capillaries.