Seitenansicht 39
30 1100 Series Capillary LC System Manual
2 Optimizing Performance
Hints for Successful Use of the Capillary Pump
Pump issues
Flush the pump extensively. First with in the purge mode, second with a
pressure applied to remove all the gas bubbles. It is recommended to do this
first with 100% A and than 100%B.
The system pressure must be higher than 20 Bar at the pump outlet.
In micro mode abnormally high column flow variations are an indication of
dirt within the system, blocked filters or loose pump valves.
Place solvent cabinet with the solvent bottles always on top (or at a higher
level) of the capillary pump.
Prevent blocking of solvent inlet filters (never use the pump without solvent
inlet filter). Growth of algae should be avoided.
When using buffer solutions, flush the system with water before switching
it off.
Check the pump plungers for scratches when changing the piston seals.
Scratched plungers will lead to micro leaks and will decrease the lifetime of
the seal.
After changing the plunger seals, perform the seal wear-in procedure. See
the pump reference manual.
Place the aqueous solvent on channel A and the organic solvent on
channel B. The default compressibility and flow sensor calibration settings
are set so. Always use the correct calibration values.
For generation of fast gradients on short columns remove the mixer, enter
the new pump configuration and select the fast gradient range for the
primary flow rate (chromatographic performance will not be affected).
When running the micro mode check the correct instrument setup (flow
sensor type, used mixer and filter).
Make sure to observe the minimum recommended flow setpoint:
Normal mode100 µl/min
Micro mode, 20 µl flow sensor: 1 µl/min
Micro mode, 100 µl flow sensor:10 µl/min