1100 Series Capillary LC System Manual 41
Optimizing Performance 2
How to Optimize the Compressibility Compensation Setting
The compressibility compensation default settings are 50 × 10
-6
/bar (best for
most aqueous solutions) for pump head A and 115 × 10
-6
/bar (to suit organic
solvents) for pump head B. The settings represent average values for aqueous
solvents (A side) and organic solvents (B side). Therefore it is always
recommended to use the aqueous solvent on the A side of the pump and the
organic solvent on the B side. Under normal conditions the default settings
reduce the pressure pulsation to values (below 1 % of system pressure) that
will be sufficient for most applications. If the compressibility values for the
solvents used differ from the default settings, it is recommended to change the
compressibility values accordingly. Compressibility settings can be optimized
by using the values for various solvents described in Table 6 on page 42. If the
solvent in use is not listed in the compressibility table, when using premixed
solvents and if the default settings are not sufficient for your application the
following procedure can be used to optimize the compressibility settings:
1 Start channel A of the capillary pump with the adequate flow rate. The
system pressure must be between 50 and 250 bar
2 Before starting the optimization procedure, the flow must be stable. Use
degassed solvent only. Check the tightness of the system with the pressure
test.
3 Your capillary pump must be connected to an Agilent ChemStation or an
Agilent 1100 control module, the pressure and%-ripple can be monitored
with one of these instruments,
otherwise connect a signal cable between the pressure output of the
capillary pump and a recording device (for example, 339X integrator) and
set parameters.
Zero 50 %
Att 2^3
Chart Speed 10 cm/min
4 Start the recording device with the plot mode.
NOTE
Use the capillary pump in the Normal Mode at least 100 µl/min.