1100 Series Capillary LC System Manual 159
Options 6
9 Insert the PEEK capillaries from the flow cell body
together with the SST fitting, ferrule and cone into
the union and tighten it.
Remove the flow cell and perform a leak test.
If no leak is observed, install the flow cell and you are
ready to work.
Make sure that the flow cell assembly is inserted
correctly and fits perfectly in the optical unit
(especially when PEEK capillaries are used).