146 1100 Series Capillary LC System Manual
6Options
Replacing Rotor Seal of Micro Column Switching Valve
15 Remove the stator head and the rotor seal.
16 Install the new rotor seal, re-install the stator head.
17 Insert the stator screws in the stator head. Tighten the screws alternately
two turns at a time until the stator head is secure.
18 Reconnect the pump capillaries to the valve ports. Slide the waste tube into
the waste holder in the leak tray.
19 Perform a pressure-tightness test to ensure the valve is pressure tight to
400 bar.
Frequency If valve leaks
Tools re quired 5.5 mm wrench
9/64 inch hex key
Parts required Refer to “500 nl Flow Cell Kit G1315-68714" on page 153.
1 Remove capillaries from ports 1, 5, and 6. 2 Loosen each fixing stator screws two turns at a
time. Remove bolts from head.