156 1100 Series Capillary LC System Manual
6Options
Installation of the Flow Cell
The flow cell is supplied with blank capillaries at the instrument side to allow
the use of different fittings, see Figure 44 on page 154.
If you are using small i.d capillary columns from e.g. LC Packings, see also
“Connecting Small I.D. Capillaries” on page 160.
The steps below describe the connection to the internal hydraulic connector
and might not be used in case the capillaries are routed directly to the column
and/or waste
.
1 Disconnect the capillaries from the capillary holder
and remove the flow cell.
2 Insert the union “Top Adjust” (supplied with the kit)
into the lower capillary holder position. This will be
used as tool for fixing the ferrules at the capillaries.