Seitenansicht 189
180 1100 Series Capillary LC System Manual
A Safety Information
High concentrations of inorganic acids like nitric acid, sulfuric acid
especially at higher temperatures (replace, if your chromatography method
allows, by phosphoric acid or phosphate buffer which are less corrosive
against stainless steel).
Halogenated solvents or mixtures which form radicals and/or acids, for
example:
2CHCl
3
+ O
2
2COCl
2
+ 2HCl
This reaction, in which stainless steel probably acts as a catalyst, occurs
quickly with dried chloroform if the drying process removes the stabilizing
alcohol.
Chromatographic grade ethers, which can contain peroxides (for example,
THF, dioxane, di-isopropylether) such ethers should be filtered through dry
aluminium oxide which adsorbs the peroxides.
Solutions of organic acids (acetic acid, formic acid, and so on) in organic
solvents. For example, a 1-% solution of acetic acid in methanol will attack
steel.
Solutions containing strong complexing agents (for example, EDTA,
ethylene diamine tetra-acetic acid).
Mixtures of carbon tetrachloride with 2-propanol or THF.
Avoid the use of alkaline solutions (pH > 8.5) which can attack the fuse
silica from the capillaries.
Agilent Technologies on Internet
For the latest information on products and services visit our worldwide web
site on the Internet at:
http://www.agilent.com
Select Products > Chemical Analysis
It will provide also the latest firmware of the Agilent 1100 series modules for
download.