32 1100 Series Capillary LC System Manual
2 Optimizing Performance
Sampler issues
For fast gradient use valve to bypass function after the sample is transferred
to the column. This function results in smaller delay times and sharper
gradient curves.
When doing automated gradient runs, use the fast composition
change/reconditioning function to equilibrate the system between runs.
Column thermostat issues
Use the column brackets to put the column in contact with the heat
exchanger.
Do not use the solvent pre-heating path (heat exchanger in the column
compartment) when you are working with capillary columns. The
dispersion will be too high.
DAD issues
At very low flow rates, bubbles might form in the cell due to low pressure in
the cell. This might cause the detector signal to have spikes and noise.
Adding a 50 µm capillary to the outlet of the cell can reduce this effect.
To avoid cell damage due to overpressure, set the upper pressure limit to
50 bar greater than the typical operating pressure.