Appendix F 751
Replacing the 8753ES with the E5070B/E5071B
Comparing Functions
F. Replacing the 8753ES
with the E5070B/E5071B
Comparing Functions
Table F-5 Functions of the 8753ES vs. the E5070B/E5071B
Function 8753ES E5070B/E5071B
Measurement Reset Can be executed by using the front
panel and the GPIB command.
Can be executed by using the front
panel, the GPIB command, and
telnet.
Channel Number of channels 4 channels 4/9/12/16 channels
Coupling between channels Channels 3 and 4 are auxiliary
channels for channels 1 and 2 and
subject to the same sweep
conditions, etc. applicable to
channels 1 and 2.
Couple/Uncouple between channels
1 and 2 can be set up.
Channels are independent of each
other.
Trace A trace for each channel Each channel can accommodate up
to 4/9/12/16 traces.
Measurement parameter S-parameters, A, B, R, A/R, B/R,
A/B, and analog bus
S-parameters (Mixed-mode
S-parameters),
Absolute value meas parameter
S-parameter conversion Impedance (reflection and
transmission), admittance (reflection
and transmission), and 1/S
←
Display format (Data format) log magnitude, linear magnitude,
phase, group delay, Smith chart,
polar format, SWR, real, imaginary
log magnitude, linear magnitude,
phase, expanded phase, positive
phase, group delay, Smith chart,
polar format, SWR, real, imaginary
Test port output Number of ports 2 ports 2 ports (Opts. 213/214)
3 ports (Opts. 313/314)
4 ports (Opts. 413/414)
Frequency 30 kHz to 3 GHz (Std.)
30 kHz to 6 GHz (Opt. 006)
300 kHz to 3 GHz (E5070B)
300 kHz to 8.5 GHz (E5071B)
Power level -85 to +10 dBm (Std.)
-85 to +8 dBm (Opts. 014/075)
-15 to 10 dBm (Opts. 213, 313, 413)
-50 to 10 dBm (Opts. 214, 314, 414)
Characteristic Impedance 50 Ω(Std.)
75 Ω(Opt. 075)
50 Ω
Coupled/Uncoupled power
levels
You can define Couple/Uncouple
between channels and between ports.
Traces are coupled on the same
channel, not between channels.
Power slope function Available ←
Turning the output On/Off Allowed ←
Power range Items to be set
up
Per port and per channel Per channel (Opts. 214/314/414)
If the above options are not installed,
the default range is -15 to 10 dBm.
Automatic
setting
On/Off setting capability Manual setting
Setting range Range setting with 10 dBm
resolution is definable from between
-15 and +10 dBm down to between
-85 and -60 dBm.
Range setting with 5 dBm resolution
is definable from between -15 and 10
dBm down to between -50 and -25
dBm