Agilent Technologies ENA Series Betriebsanweisung Seite 113

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 775
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 112
Chapter 4 107
Calibration
OPEN/SHORT Response Calibration (reflection test)
4. Calibration
OPEN/SHORT Response Calibration (reflection test)
In OPEN or SHORT response calibration, calibration data are measured by connecting an
OPEN or SHORT standard, respectively, to the desired test port. For frequency response,
these calibrations effectively eliminate the reflection tracking error from the test setup in a
reflection test using that port (Figure 4-3). It is also possible to carry out isolation
calibration with a LOAD standard during OPEN/SHORT response calibration. An
isolation calibration will eliminate the directivity error from the test setup in a reflection
test using that port (Figure 4-4).
Figure 4-3 1-Port error model (OPEN/SHORT response)
Figure 4-4 1-Port error model (OPEN/SHORT response + isolation)
Procedure
Step 1. Press or to select the channel for which you want to perform the
calibration.
Step 2. Press .
Step 3. Press
Calibrate.
Seitenansicht 112
1 2 ... 108 109 110 111 112 113 114 115 116 117 118 ... 774 775

Kommentare zu diesen Handbüchern

Keine Kommentare