Chapter 11 389
Limit Test
Ripple Test
11. Limit Test
Figure 11-10 Example of ripple limit configuration (for limit table in Figure 11-9)
NOTE The individual frequency bands for the ripple test can overlap each other; in this case, the
ripple limit test is performed for each frequency band.
NOTE Even if the E5070B/E5071B's span value is set to zero, you must enter a parameter for both
Begin Stimulus and End Stimulus.
NOTE If the data format is Smith chart or polar, the limit test is performed for the main response
value among the two marker response values.
Saving/recalling ripple limit table
The ripple limit table can be saved in a file and recalled later for use on the screen. The file
is saved in the csv format (with the extension *.csv), and values are saved as a character
string with the unit. The csv formatted file can also be reused in spreadsheet software made
for PCs.
The ripple limit table is saved in the following format:
Example 11-2 Sample Ripple Limit Table Saved in CSV Format
"# Channel 1"
"# Trace 1"
Type, Begin Stimulus, End Stimulus, MaxRipple
ON, 933.0000000 MHz, 964.0000000 MHz, 1.5 dB
ON, 938.0000000 MHz, 953.0000000 MHz, 500 mdB
ON, 953.0000000 MHz, 960.0000000 MHz, 300 mdB