Chapter 4 173
Calibration
User-characterized ECal
4. Calibration
5. Measuring User Characteristics
Step 1. Select
Characterize ECAL (1 in Figure 4-47) to display the User Characteristic
Measurement screen.
Figure 4-47 EcalCharacterization macro (User Characteristic Measurement screen)
Step 2. After connecting the adapter to the ECal module as necessary, connect each port of the
ECal module and the test port of the E5070B/E5071B.
Figure 4-48 Connecting ECal module (for 4 ports)
NOTE You can select any port of the ECal module and any test port of the E5070B/E5071B for
connection; the E5070B/E5071B automatically recognizes the connected ports before
measurement.
Step 3. Press
Measure (2 in Figure 4-47) to start measurement.