Chapter 14 481
Controlling E5091A
Setting the E5091A
14. Controlling E5091A
Selecting the E5091A Model
Select the E5091A model you want to set. For the E5091A-016, you can select the 13-port
device or 16-port device function.
Step 1. Press -
E5091A Setup to display the E5091A setup menu.
Step 2. After the execution of “Selecting ID for E5091A” on page 480, select the model by
pressing
Select Test Set.
NOTE If the model you use and the selected softkey is different, the configuration will not be
reflected. Also, no error message will appear. For the correlation between the model and
the softkey, refer to the Table 14-2.
Assigning test ports
Before calibration and measurement, you need to assign the test ports of the E5091A. You
can set the connection ports for each channel and perform measurement while switching
the connection for each channel.
Operational procedure
Step 1. Press -
Multiport Test Set Setup to display the E5091A setup menu.
Step 2. Press (or ) to activate the channel for which you want to set the
connection ports.
Step 3. Use the corresponding softkey to assign the test ports for the connection between the
E5091A and the interconnection ports.
NOTE The port name of the softkey indicates the E5091A interconnection port.
Softkey Model
E5091_9 Select the E5091A option 009.
E5091_13 Select the E5091A option 016 for the 13-port device function.
E5091_16 Select the E5091A option 016 for the 16-port device function.
Table 14-2 The Model you use and the softkey you select
E5091_009 E5091_013 E5091_016
E5091A-009 Yes No No
E5091A-016
No
Yes Yes
Table 14-3 When the E5091A-009 is Connected
Softkey Function
Port1 Selects a test port of the E5091A to which you want to connect port 1 of the
E5091A. You can select the port from
A or T1
*1
.