Chapter 16 541
Specifications and Supplemental Information
Corrected System Performance
16. Specifications and
Supplemental Information
Table 16-3 Corrected System Performance With Type-N Device Connectors, 85092C
Electronic Calibration Module
Network analyzer: E5070B/E5071B, Calibration module: 85092C (Type-N, 50 Ω) electronic calibration (ECal)
module, Calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23°C±5°C with <1°C deviation
from calibration temperature, Isolation calibration not omitted
Description
Specification (dB)
3 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz
Directivity 52 52 47
Source Match 45 41 36
Load Match 47 44 39
Reflection Tracking ±0.040 ±0.060 ±0.070
Transmission Tracking ±0.039 ±0.069 ±0.136
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
Kommentare zu diesen Handbüchern