480 Chapter 14
Controlling E5091A
Setting the E5091A
Setting the E5091A
This section describes the settings of the E5091A. Table 14-1 shows the flow used for item
setting.
Selecting ID for E5091A
Set the target ID to the ID of the connected E5091A.
Step 1. Press -
Multiport Test Set Setup to display the E5091A setup menu.
Step 2. Select
Test Set 1 for ID 1 and Test Set 2 for ID 2. The ID is set with the bit switch on the
rear panel of the E5091A (Figure 14-3).
NOTE Change the ID bit switch setting while the E5070B/E5071B is turned off.
Figure 14-3 ID bit switch of E5091A
Table 14-1 Setting flow for E5091A
Item Description
“Selecting ID for E5091A” on page 480 Selects the ID of the E5091A you want to
set
“Selecting the E5091A Model” on
page 481
Selects the E5091A model you want to set
“Assigning test ports” on page 481 Assigns test ports of the E5070B/E5071B
and those of the E5091A
“Displaying the E5091A properties” on
page 483
Displays the E5091A property to check the
port setting
“Setting control line” on page 484 Makes the setting of the control line that
controls the DUT
“Enabling control of E5091A” on
page 485
Enables functions to control the E5091A