Agilent Technologies ENA Series Betriebsanweisung Seite 151

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 775
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 150
Chapter 4 145
Calibration
4-port TRL calibration
4. Calibration
Figure 4-35 Connection of standards in 4-port TRL calibration
Operational procedure
NOTE The shapes of actual standards differ from those shown in Figure 4-35 which are just
symbols for illustration.
Step 1. Press or to select the channel for which you want to execute
calibration.
Step 2. Press .
Step 3. Press
Calibrate.
Step 4. Press
4-Port TRL Cal.
Step 5. Press
Select Ports.
Step 6.
Thru/Line to start the measurement of the calibration standard.
Softkey Function
Port 1-2 Thru Executes thru/line calibration for test ports 1 and 2.
Port 1-3 Thru Executes thru/line calibration for test ports 1 and 3.
Port 1-4 Thru Executes thru/line calibration for test ports 1 and 4.
Port 2-3 Thru Executes thru/line calibration for test ports 2 and 3.
Port 2-4 Thru Executes thru/line calibration for test ports 2 and 4.
Port 3-4 Thru Executes thru/line calibration for test ports 3 and 4.
Seitenansicht 150
1 2 ... 146 147 148 149 150 151 152 153 154 155 156 ... 774 775

Kommentare zu diesen Handbüchern

Keine Kommentare