Agilent Technologies ENA Series Betriebsanweisung Seite 210

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 775
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 209
204 Chapter 4
Calibration
Specifying Different Standard for Each Frequency
Disabling standard defined for a subclass
The following procedure shows how to disable a standard defined for a subclass. Note that
you cannot disable subclass 1 because at least one standard must exist. The following
procedure shows how to disable an OPEN standard of subclass 2.
Procedure
Step 1. Press .
Step 2. Press
Cal Kit.
Step 3. Select a calibration kit you want to use. (In this example, use the calibration kit defined in
Step 3 of “Defining different standard for each frequency band” on page 202.)
Step 4. Press
Modify Kit.
Step 5. Press
Specify CLSs.
Step 6. Press
Sub Class, and select Sub Class 2.
Step 7. Select the standard type
Open, press Set All, make settings for all ports.
Step 8. Select
None.
Step 9. Press
Return.
Seitenansicht 209
1 2 ... 205 206 207 208 209 210 211 212 213 214 215 ... 774 775

Kommentare zu diesen Handbüchern

Keine Kommentare