Obsolete and Discontinued Commands 7
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 705
:CHANnel:ACTivity
(see page 798)
Command Syntax
:CHANnel:ACTivity
The :CHANnel:ACTivity command clears the cumulative edge variables for
the next activity query.
Query Syntax
:CHANnel:ACTivity?
The :CHANnel:ACTivity? query returns the active edges since the last clear,
and returns the current logic levels.
Return Format
<edges>,<levels><NL>
<edges> ::= presence of edges (32-bit integer in NR1 format).
<levels> ::= logical highs or lows (32-bit integer in NR1 format).
A bit equal to one indicates that edges have been detected at the specified
threshold since the last clear on that channel.
NOTE
The :CHANnel:ACTivity command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :ACTivity command (see page 154) instead.
NOTE
A bit equal to zero indicates that no edges were detected at the specified threshold since
the last clear on that channel. Edges may have occurred that were not detected because of
the threshold setting.