Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 493
:TRIGger:CAN:PATTern:DATA:LENGth
(see page 798)
Command Syntax
:TRIGger:CAN:PATTern:DATA:LENGth <length>
<length> ::= integer from 1 to 8 in NR1 format
The :TRIGger:CAN:PATTern:DATA:LENGth command sets the number of
8- bit bytes in the CAN data string. The number of bytes in the string can
be anywhere from 0 bytes to 8 bytes (64 bits). The value for these bytes is
set by the :TRIGger:CAN:PATTern:DATA command.
Query Syntax
:TRIGger:CAN:PATTern:DATA:LENGth?
The :TRIGger:CAN:PATTern:DATA:LENGth? query returns the current CAN
data pattern length setting.
Return Format
<count><NL>
<count> ::= integer from 1 to 8 in NR1 format
Errors "- 241, Hardware missing" on page 757