Agilent Technologies InfiniiVision 6000 Series Wartungshandbuch Seite 576

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 934
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 575
576 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:LIN:SYNCbreak
(see page 798)
Command Syntax
:TRIGger:LIN:SYNCbreak <value>
<value> ::= integer = {11 | 12 | 13}
The :TRIGger:LIN:SYNCbreak command sets the length of the LIN sync
break to be greater than or equal to 11, 12, or 13 clock lengths. The sync
break is the idle period in the bus activity at the beginning of each packet
that distinguishes one information packet from the previous one.
Query Syntax
:TRIGger:LIN:SYNCbreak?
The :TRIGger:LIN:STANdard? query returns the current LIN sync break
setting.
Return Format
<value><NL>
<value> ::= {11 | 12 | 13}
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:MODE" on page 485
":TRIGger:LIN:SIGNal:DEFinition" on page 752
":TRIGger:LIN:SOURce" on page 574