368 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:MEASure:XMAX
(see page 798)
Command Syntax
:MEASure:XMAX [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:XMAX command installs a screen measurement and starts
an X-at- Max-Y measurement on the selected window. If the optional
source parameter is specified, the current source is modified.
Query Syntax
:MEASure:XMAX? [<source>]
The :MEASure:XMAX? query measures and returns the horizontal axis
value at which the maximum vertical value occurs. If the optional source
is specified, the current source is modified. If all channels are off, the
query returns 9.9E+37.
Return Format
<value><NL>
<value> ::= horizontal value of the maximum in NR3 format
See Also "Introduction to :MEASure Commands" on page 324
":MEASure:XMIN" on page 369
":MEASure:TMAX" on page 729
NOTE
:MEASure:XMAX is an alias for :MEASure:TMAX.