Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 449
:SBUS:SPI:BITorder
(see page 798)
Command Syntax
:SBUS:SPI:BITorder <order>
<order> ::= {LSBFirst | MSBFirst}
The :SBUS:SPI:BITorder command selects the bit order, most significant bit
first (MSB) or least significant bit first (LSB), used when displaying data
in the serial decode waveform and in the Lister.
Query Syntax
:SBUS:SPI:BITorder?
The :SBUS:SPI:BITorder? query returns the current SPI decode bit order.
Return Format
<order><NL>
<order> ::= {LSBF | MSBF}
Errors "- 241, Hardware missing" on page 757
See Also "Introduction to :SBUS Commands" on page 433
":SBUS:MODE" on page 448
":TRIGger:SPI Commands" on page 593
NOTE
This command is only valid on 4 (analog) channel oscilloscope models when the low-speed
IIC and SPI serial decode option (Option LSS) has been licensed.