444 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:SBUS:I2S:BASE
(see page 798)
Command Syntax
:SBUS:I2S:BASE <base>
<base> ::= {DECimal | HEX}
The :SBUS:I2S:BASE command determines the base to use for the I2S
decode display.
Query Syntax
:SBUS:I2S:BASE?
The :SBUS:I2S:BASE? query returns the current I2S display decode base.
Return Format
<base><NL>
<base> ::= {DECimal | HEX}
Errors "- 241, Hardware missing" on page 757
See Also "Introduction to :SBUS Commands" on page 433
":TRIGger:I2S Commands" on page 538
NOTE
This command is only valid on 4 (analog) channel oscilloscope models when the I2S serial
decode option (Option SND) has been licensed.