Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 201
:ACQuire:RSIGnal
(see page 798)
Command Syntax
:ACQuire:RSIGnal <ref_signal_mode>
<ref_signal_mode> ::= {OFF | OUT | IN}
The :ACQuire:RSIGnal command selects the 10 MHz reference signal mode.
The OFF mode disables the oscilloscope's 10 MHz REF BNC connector.
The OUT mode is used to synchronize the timebase of two or more
instruments.
The IN mode is used to supply a sample clock to the oscilloscope. A
10 MHz square or sine wave signal is input to the BNC connector
labeled 10 MHz REF. The amplitude must be between 180 mV and 1 V,
with an offset of between 0 V and 2 V.
Query Syntax
:ACQuire:RSIGnal?
The :ACQuire:RSIGnal? query returns the current 10 MHz reference signal
mode.
Return Format
<ref_signal_mode><NL>
<ref_signal_mode> ::= {OFF | OUT | IN}
See Also ":TIMebase:REFClock" on page 472
The Agilent InfiniiVision 6000 Series Oscilloscope User's Guide for
information on using the 10 MHz reference clock.
CAUTION
Do not apply more than ±15 V at the 10 MHz REF BNC connector on the rear panel,
or damage to the instrument may occur.