What's New 1
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 43
Command Differences From 54620/54640 Series Oscilloscopes
The main differences between the version 1.00 programming command set
for the InfiniiVision 6000 Series oscilloscopes and the 54620/54640 Series
oscilloscopes are related to:
• :HARDcopy and :DISPlay command subsystem changes for USB printers
and the high resolution color display.
• New standards supported by the :TRIGger:TV commands.
• Support for 113xA Series probes.
• New "RAW" :WAVeform:POINts option for retrieving raw acquisition
record data.
• Discontinuance of the common commands for macros.
More detailed descriptions of the new, changed, obsolete, and discontinued
commands appear below.
New Commands
Command Description
:ACQuire:RSIGnal (see page 201) Selects the 10 MHz reference signal mode.
:CHANnel<n>:PROBe:ID? (see page 241) Returns the type of probe attached to the
specified oscilloscope channel.
:CHANnel<n>:PROBe:STYPe (see page 243) Sets the channel probe signal type (STYPe) to
differential or single-ended when using the
113xA Series probes, and determines how
offset is applied.
:CHANnel<n>:VERNier (see page 248) Specifies whether the channel's vernier (fine
vertical adjustment) setting is ON (1) or OFF (0).
:DIGital<n>:SIZE (see page 254) Specifies the size of digital channels on the
display.
:EXTernal:PROBe:ID (see page 271) Returns the type of probe attached to the
external trigger input.
:EXTernal:PROBe:STYPe (see page 272) Sets the external trigger probe signal type
(STYPe) to differential or single-ended when
using the 113xA Series probes, and determines
how offset is applied.
:HARDcopy:FILename (see page 720) Sets the output filename for print formats
whose output is a file. Replaces the 5462x/4x
:HARDcopy:DESTination (see page 718)
command.
:HARDcopy:PDRiver (see page 724) Sets the hardcopy printer driver.
:HARDcopy:IGColors (see page 723) Specifies whether graticule colors are inverted.
Kommentare zu diesen Handbüchern