Installing your System
Flow Connections to the VW Detector
Flow Connections to the VW Detector
WARNING When opening capillary or tube fittings, solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used.
NOTE The flow cell is shipped with a filling of isopropanol (also recommended when
the instrument and/or flow cell is shipped to another location). This is to avoid
breakage due to subambient conditions.
1 Press the release buttons and remove the front cover to have access to the
flow cell area.
2 Remove the metal cover from the flow cell area and install the flow cell.
Tighten the cell screws (see Figure 11 on page 29).
3 Connect the column-detector capillary to the flow cell inlet connector.
4 Connect the other end of the column-detector capillary to the column.
5 Connect the PEEK waste capillary to the flow cell outlet connector (the PEEK
capillary introduces a small back pressure to the flow cell for better baseline
stability).
6 Connect the corrugated waste tube to the waste outlet of the detector leak
handling system.
7 Route the waste tubing to your waste container.
Preparations Detector is installed in the HPLC system.
Parts required Other modules
Parts from accessory kit G1314-68705
Two wrenches 1/45/16 inch for capillary connections
Seitenansicht 27