Installing your System
Flow Connections at the Manual Injector
Flow Connections at the Manual Injector
WARNING When opening capillary or tube fittings, solvents may leak out. Please
observe appropriate safety procedures (for example, goggles, safety
gloves and protective clothing) as described in the material handling
and safety data sheet supplied by the solvent vendor, especially when
toxic or hazardous solvents are used (see also Leak Drainage of the
Manual Injector on page 25).
1 Connect the pump outlet capillary to port 2.
2 Connect the column outlet capillary to port 3.
3 Connect the sample loop between ports 1 and 4.
4 Connect one vent capillary (supplied with valve) to port 5, and one to port 6.
CAUTION The outlets of the two vent capillaries (ports 5 and 6) and the needle port must
be at the same level to prevent siphoning (see Figure 8).
Figure 7 Flow Connections
Sample loop
Waste
Vent tube
Sample loop
To column
From pump outlet