Agilent Technologies E5071C Betriebsanweisung Seite 65

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 551
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 64
Chapter 4 65
Performing a Calibration
Calibration
4. Performing a Calibration
As shown in Table 4-2, you need to set different items depending on the standard type.
You need to set the items identified by marks in the table above.
Defining a Standard Class Assignment
To select the standard to be applied to the OPEN measurement for each port, use the
following command:
:SENS{1-4}:CORR:COLL:CKIT:ORD:OPEN on page 378
To select the standard to be applied to the SHORT measurement for each port, use the
following command:
:SENS{1-4}:CORR:COLL:CKIT:ORD:SHOR on page 379
To select the standard to be applied to the LOAD measurement for each port, use the
following command:
:SENS{1-4}:CORR:COLL:CKIT:ORD:LOAD on page 377
To select the standard to be applied to the THRU measurement between each pair of ports,
use the following command:
:SENS{1-4}:CORR:COLL:CKIT:ORD:THRU on page 380
Table 4-2 Settings Specific to Each Standard Type
Standard Types C0 to C3 L0 to L3
Offset
Delay
Offset
Loss
Offset Z0
Arbitrary
Impedance
OPEN √√√
SHORT √√√√
LOAD √√√
THRU √√√
Arbitrary
Impedance
√√√
Seitenansicht 64
1 2 ... 60 61 62 63 64 65 66 67 68 69 70 ... 550 551

Kommentare zu diesen Handbüchern

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm